Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa 7544-02: (1 year) CAL VER /5520 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 33 NUMBER OF LINES: 144 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 1 hour - REQUIRED 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {AC VOLTAGE} 1.008 DISP Press the POWER button to turn the unit ON 1.008 DISP 1.008 DISP Set the Function Switch to AC VOLTS. 1.009 DISP Connect the 5520A to the UUT as follows: 1.009 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.009 DISP [32] 5520A NORMAL LO to UUT COM 1.010 HEAD AC VOLTAGE: {500mV Range} 1.011 5520 500 400.00mV 0.7% 0.3U 1kH SI 2W 2.001 5520 500 400.00mV 2% 0.5U 20kH SI 2W 3.001 HEAD AC VOLTAGE: {5V Range} 3.002 5520 5 4.0000V 1% 0.003U 30H SI 2W 4.001 5520 5 4.0000V 0.5% 0.003U 10kH SI 2W 5.001 5520 5 4.0000V 1% 0.004U 20kH SI 2W 6.001 5520 5 4.0000V 2% 0.007U 50kH SI 2W 7.001 5520 5 4.0000V 3% 0.02U 100kH SI 2W 8.001 HEAD AC VOLTAGE: {50V Range} 8.002 5520 50 40.000V 0.5% 0.03U 1kH SI 2W 9.001 HEAD AC VOLTAGE: {500V Range} 9.002 5520 500 400.00V 0.5% 0.3U 1kH SI 2W 10.001 HEAD AC VOLTAGE: {750V Range} 10.002 5520 750 700.0V 1% 3U 100H SI 2W 11.001 5520 * S 11.002 HEAD {FREQUENCY} 11.003 DISP Press the SELECT button as needed 11.003 DISP to select the HZ function. 11.004 5520 9.000kH 0.05% 0.001U 1V SI 2W 12.001 HEAD {dBm} 12.002 DISP Press the SELECT button as needed 12.002 DISP to select the dBm function. 12.003 5520 0.7746V 1kH SI S 2W 12.004 ACC 0.0dBm 0.01U 12.005 MEMI Enter the reading in dBm 12.006 MEME 12.007 MEMC dBm 0.1U 13.001 HEAD {DC VOLTAGE} 13.002 DISP Set the Function Switch to DC VOLTS. 13.003 HEAD DC VOLTAGE: {5V Range} 13.004 5520 5 4.0000V 0.06% 0.0002U 2W 14.001 HEAD DC VOLTAGE: {50V Range} 14.002 5520 50 40.000V 0.06% 0.002U 2W 15.001 HEAD DC VOLTAGE: {500V Range} 15.002 5520 500 400.00V 0.06% 0.02U 2W 16.001 HEAD DC VOLTAGE: {1000V Range} 16.002 5520 1000 1000.0V 0.06% 0.2U 2W 17.001 5520 * S 17.002 HEAD DC VOLTAGE: {500mV Range} 17.003 DISP Set the Function Switch to 500mV DC 17.004 ASK- R 17.005 5520 400.00mV 0.04% 0.02U 2W 18.001 ASK+ R 18.002 HEAD {DIODE TEST} 18.003 DISP Set the Function Switch to DIODE. 18.004 5520 0.5000V 1% 0.0002U 2W 19.001 HEAD {RESISTANCE} 19.002 DISP Set the Function Switch to OHMS 19.003 HEAD RESISTANCE: {50 MOhms} 19.004 5520 50 10.000MZ 1% 0.004/ 2W 20.001 HEAD RESISTANCE: {5 MOhms} 20.002 5520 5 1.0000MZ 0.5% 0.004/ 2W 21.001 HEAD RESISTANCE: {500 kOhms} 21.002 DISP Connect the 5520A to the UUT as follows: 21.002 DISP [32] 5520A AUX HI to UUT V/OHM 21.002 DISP [32] 5520A AUX LO to UUT COM 21.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 21.002 DISP [32] 5520A NORMAL LO to UUT COM 21.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 21.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 21.003 5520 500 100.00kZ 0.06% 0.004/ CW 22.001 HEAD RESISTANCE: {50 kOhms} 22.002 5520 50 10.000kZ 0.06% 0.004/ CW 23.001 HEAD RESISTANCE: {5 kOhms} 23.002 5520 5 1.0000kZ 0.06% 0.004/ CW 24.001 HEAD RESISTANCE: {500 Ohms} 24.002 5520 500 100.00Z 0.06% 0.004/ CW 25.001 ASK- R 25.002 HEAD {500uA} 25.003 DISP Set the Function Switch to 500uA. 25.004 DISP Connect the 5520A to the UUT as follows: 25.004 DISP [32] 5520A AUX HI to UUT mA/uA 25.004 DISP [32] 5520A AUX LO to UUT COM 25.005 5520 500 400.00uA 0.3% 0.004/ 2W 26.001 DISP Press the SELECT button to select AC current. 26.002 5520 500 400.00uA 1% 0.04/ 100H SI 2W 27.001 HEAD {50mA} 27.002 DISP Set the Function Switch to 50mA. 27.003 5520 50 40.000mA 0.3% 0.004/ 2W 28.001 DISP Press the SELECT button to select AC current. 28.002 5520 50 40.000mA 1% 0.04/ 100H SI 2W 29.001 HEAD {500mA} 29.002 DISP Set the Function Switch to 500mA. 29.003 5520 500 400.00mA 0.3% 0.004/ 2W 30.001 DISP Press the SELECT button to select AC current. 30.002 5520 500 400.00mA 1% 0.04/ 100H SI 2W 31.001 HEAD {10A} 31.002 DISP Set the Function Switch to 10A. 31.003 DISP Connect the 5520A to the UUT as follows: 31.003 DISP [32] 5520A AUX HI to UUT 10A 31.003 DISP [32] 5520A AUX LO to UUT COM 31.004 5520 10 2.000A 0.7% 0.2/ 2W 32.001 DISP Press the SELECT button to select AC current. 32.002 5520 10 2.000A 1% 0.2/ 100H SI 2W 33.001 END